کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1884251 1043334 2011 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A pixel detector asic for dosimetry using time-over-threshold energy measurements
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
A pixel detector asic for dosimetry using time-over-threshold energy measurements
چکیده انگلیسی

In this work we present the design of a chip which provides the readout of a highly segmented diode array, in which signals induced by individual X-ray photons are processed discretely. There are several benefits to this approach, including the ability to achieve a high signal to noise ratio due to the inherently low sensor capacitance, and the suppression of background noise (e.g. dark current) using an analogue threshold. The segmentation also ensures a linear behaviour even at very high dose rates. A time over threshold (ToT 1) energy measurement technique provides an immediate digital value corresponding to the energy deposited onto the diode by each individual photon. Deadtime-free operation is achieved by reading out a subset of the detector segments at a time while the rest of the detector continues to process signals. This paper describes the application-specific integrated circuit (ASIC) chip which was designed to provide pre-processing of photo-induced signals in the detector and readout of the processed digital data.


► We have designed the ASIC of a hybrid pixel detector for photon dosimetry.
► The pulse processing circuits measure the time over threshold of discrete photons.
► Each pixel automatically sorts ToT measurements into 16 digital energy bins.
► The binned data is used to reconstruct the charge spectrum and dose in real time.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Measurements - Volume 46, Issue 12, December 2011, Pages 1619–1623
نویسندگان
, , , , , , , , , , , , , ,