Study of the filter method for neutron pulse-height distributions measured with organic scintillators
ترجمه فارسی عنوان
بررسی روش فیلتر برای توزیع ارتفاع پالس نوترونی با اندازه گیری اسکنرهای آلی
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی


• Organic scintillators sensitive to both neutrons and gamma-rays.
• The filter method includes two measurements: with and without polyethylene filter.
• Digital pulse-shape discrimination method was used as the reference method.
• The filter method is less accurate than the digital PSD method.
• The filter method has potential for robust measurement.

Identification of neutron sources is of interest in several fields, such as homeland security or basic nuclear physics. Organic scintillators offer the possibility to unfold fast-neutron energy spectra. Since organic scintillators are sensitive to neutrons and gamma-rays, a method to discriminate these two types of particles is required. This paper is focused on investigation of the discrimination properties of the filter method by using a new procedure. We compare the results of an experimental filter method to a digital pulse-shape discrimination (PSD) method based on charge integration. In addition, these methods are compared to the simulation results obtained with the MCNPX-PoliMi code. The experimental and numerical investigations were performed with a 252Cf spontaneous-fission neutron source. The integration of the number of counts gave the relative differences between the experimental filter and digital PSD neutron pulse-height distributions (PHDs) and simulated PHDs less than approximately 5% in the range between 60 keVee and 1.715 MeVee of light output. Above 1.715 MeVee, the PSD method has advantages over the filter method, due to the filter method having significantly worse counting statistics. The results show that the filter method has potential for robust neutron measurements when the PSD method cannot be applied, such as for ‘old’ organic plastic scintillators without PSD capability.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Measurements - Volume 86, March 2016, Pages 32–38
نویسندگان
, , , , , ,