کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1885287 1043402 2010 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Experimental characterization of the IRSN long counter for the determination of the neutron fluence reference values at the AMANDE facility
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Experimental characterization of the IRSN long counter for the determination of the neutron fluence reference values at the AMANDE facility
چکیده انگلیسی

The neutron metrology and dosimetry laboratory has developed a long counter, on the basis of optimization work using Monte-Carlo simulations. The calibration of this new device was performed at the IRSN standard radionuclide source facility. Measurements were also realized at the monoenergetic neutron fields facility, AMANDE, to determine the dead time correction, the effective center and efficiency as a function of neutron energy. A good agreement was observed between the long counter results and the monoenergetic neutron fluence values determined with proton recoil counters. The results of the experimental characterization of the IRSN long counter are presented in this paper.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Measurements - Volume 45, Issue 10, December 2010, Pages 1254–1257
نویسندگان
, ,