Electron beam energy monitoring using thermoluminescent dosimeters and electron back scattering
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Electron beam energy monitoring using thermoluminescent dosimeters and electron back scattering
چکیده انگلیسی

Periodic checks of megavoltage electron beam quality are a fundamental requirement in ensuring accurate radiotherapy treatment delivery. In the present work, thermoluminescent dosimeters (TLDs) positioned on either side of a lead sheet at the surface of a water equivalent phantom were used to monitor electron beam quality using the electron backscattering method. TLD100 and TLD100H were evaluated as upstream detectors and TLD200, TLD400 and TLD500 were evaluated as downstream detectors. The evaluation assessed the test sensitivity and correlation, long and short term reproducibility, dose dependence and glow curve features. A prototype of an in-air jig suitable for use in postal TLD dose audits was also developed and an initial evaluation performed. The results indicate that the TLD100-TLD200 combination provides a sensitive and reproducible method to monitor electron beam quality. The light weight and easily fabricated in-air jig was found to produce acceptable results and has the potential to be used by radiation monitoring agencies to carry out TLD postal quality assurance audits, similar to audits presently being conducted for photon beams.


► Monitoring electron beam quality via electron backscattering was investigated.
► Different thermoluminescent materials were evaluated as detectors.
► A TLD100-TLD200 combination produced the most sensitive and reproducible results.
► An in-air jig was evaluated to allow measurements via postal dose audits.