• A Ce3+ doped SiO2 scintillation detector was applied to 192Ir HDR brachytherapy.
• Detector response resulted to be linear with the delivered dose.
• No evidence of energy and dose rate dependence resulted from the study.
Aim of this work was to study the application of a new miniaturized Ce3+ doped SiO2 scintillation detector to in vivo dosimetry in high dose rate brachytherapy. Energy, dose-rate, temperature and angular dependences of the detector response to 192Ir HDR brachytherapy fields were investigated, as well as sensitivity reproducibility and linearity. To this aim, two ad hoc phantoms were designed and developed to perform measurements in water. Intra-session reproducibility resulted to be very high, however inter-session reproducibility showed too high statistical variation. Detector response resulted to increase linearly with dose (R2 = 0.997), with no evidence of energy and dose-rate dependence. Sensitivity resulted to increase linearly with temperature (R2 = 0.995), with a 0.2% increase each °C. Finally, no significant angular dependence for the source moving around a circle in the azimuthal plane centered at the scintillator was observed. The obtained results show that the proposed detector is suitable for in vivo real-time dosimetry in high dose rate brachytherapy.
Journal: Radiation Measurements - Volume 56, September 2013, Pages 312–315