کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1921161 | 1048762 | 2010 | 5 صفحه PDF | دانلود رایگان |
![عکس صفحه اول مقاله: The safety of transcranial magnetic stimulation with deep brain stimulation instruments The safety of transcranial magnetic stimulation with deep brain stimulation instruments](/preview/png/1921161.png)
ObjectivesTranscranial magnetic stimulation (TMS) has been employed in patients with an implanted deep brain stimulation (DBS) device. We investigated the safety of TMS using simulation models with an implanted DBS device.MethodsThe DBS lead was inserted into plastic phantoms filled with dilute gelatin showing impedance similar to that of human brain. TMS was performed with three different types of magnetic coil. During TMS (1) electrode movement, (2) temperature change around the lead, and (3) TMS-induced current in various situations were observed. The amplitude and area of each evoked current were measured to calculate charge density of the evoked current.ResultsThere was no movement or temperature increase during 0.2 Hz repetitive TMS with 100% stimulus intensity for 1 h. The size of evoked current linearly increased with TMS intensity. The maximum charge density exceeded the safety limit of 30 μC/cm2/phase during stimulation above the loops of the lead with intensity over 50% using a figure-eight coil.ConclusionsStrong TMS on the looped DBS leads should not be administered to avoid electrical tissue injury. Subcutaneous lead position should be paid enough attention for forthcoming situations during surgery.
Journal: Parkinsonism & Related Disorders - Volume 16, Issue 2, February 2010, Pages 127–131