کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
229500 | 465030 | 2014 | 8 صفحه PDF | دانلود رایگان |

A new liquid chromatographic method has been developed and validated for the determination of terbutaline sulfate (TLS), guaifenesin (GFN) and ambroxol HCl (AML), for its potential impurities in drug substances and drug products. Efficient chromatographic separation was achieved on X-Terra RP-18 column with a simple mobile phase combination containing a gradient mixture of solvents A and B at a flow rate of 1.0 mL min−1 and quantitation was carried out using ultraviolet detection at 222 nm with column temperature of 35 °C. The resolution between TLS, GFN and AML, its associated impurities was found to be greater than 1.5. Regression analysis shows an r value (correlation coefficient) greater than 0.998. This method was capable to detect all the process impurities of TLS, GFN and AML, at a level below 0.015% with respect to a test concentration of 0.125, 5.0 and 1.5 mg mL−1, respectively. The % RSD for the inter-day and intra-day precisions for all the impurities of TLS, GFN and AML were found to be less than 3.0. The method has shown good, consistent recoveries. The drugs were subjected to stress conditions of acid, base, water hydrolysis, oxidation, photolysis and thermal degradation, as prescribed by international conference on harmonization (ICH).
Journal: Journal of Saudi Chemical Society - Volume 18, Issue 5, November 2014, Pages 593–600