A Comparative Study of Two Nebulizers in the Emergency Department: Breath-Actuated Nebulizer and Handheld Nebulizer
ترجمه فارسی عنوان
مطالعه مقایسه ای دو نبولایزر در بخش اورژانس: نبولایزر فعال با نفس و نبولایزر دستی
کلمات کلیدی
دستورالعمل پرستاری پیشرفته، نفس نبولایزر را فعال می کند، نبولایزر دستی، بخش اورژانس، پرستاری اضطراری، مقیاس اصلاح شده بورگ، جریان پیک
موضوعات مرتبط
علوم پزشکی و سلامت پرستاری و مشاغل بهداشتی امداد اورژانس
چکیده انگلیسی

IntroductionThe breath-actuated nebulizer (BAN) and the handheld nebulizer (HHN) are 2 nebulizers used in the ED of Cooper University Hospital. The purpose of this study was to compare the nebulizers to identify which device resulted in a resolution of symptoms with fewer treatments. The primary hypothesis was that adult ED patients with a chief complaint of wheezing and dyspnea who were given nebulized treatments via the BAN would require less nebulizer treatments than those patients given nebulized treatments via HHN. In addition, the secondary purposes of the study was to determine if the BAN would have significantly higher peak expiratory flow measurements, lower Modified Borg Score, overall decreased respiratory rate, and lower heart rates compared to subjects receiving nebulized treatments via HHN.MethodsA single-site, prospective, randomized, comparative design study was conducted in the ED between March 2010 and February 2011. Fifty-four subjects presenting with dyspnea and wheezing and an Emergency Severity Index of 3 or 4 were enrolled and randomly assigned to 1 of 2 groups (BAN or HHN). Subjects were administered 1 to 3 nebulizer treatments (#1 ipratropium bromide and albuterol sulfate, #2 ipratropium bromide and albuterol sulfate, #3 albuterol sulfate), which was consistent with the ED Advanced Nursing Guideline for Wheezing. Nebulizer treatments were discontinued if a patient’s dyspnea or wheezing resolved. IRB approval was obtained prior to study commencement.ResultsThere was no significant difference found between the HHN and BAN in respect to number of treatments, respiratory rate, peak flow measurements, and Modified Borg scores in the 54 subjects. There was a difference of 7 points in pulse rate between the pre- and post-second BAN treatment (n = 51, P = 0.01). Subjects in the BAN group who completed all 3 treatments (n = 18) had a total treatment time that was on average of 10 minutes longer than those in the HHN group.ConclusionsThis study demonstrated no clinical difference between the BAN and HHN in terms of respiratory rate, peak flow, perception of dyspnea, and number of treatments. It is possible that the longer treatment times account for the elevated pulse rate. The data suggests that the higher cost and the longer treatment time do not justify the use of the BAN in this setting. We recommend that these devices be tested with a larger sample size to further test the differences between these 2 devices.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Emergency Nursing - Volume 40, Issue 2, March 2014, Pages 131–137
نویسندگان
, , , , , , , ,