کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2642587 1137811 2012 11 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Uphill Grime: Process Improvement in Surgical Instrument Cleaning
موضوعات مرتبط
علوم پزشکی و سلامت پرستاری و مشاغل بهداشتی پرستاری
پیش نمایش صفحه اول مقاله
Uphill Grime: Process Improvement in Surgical Instrument Cleaning
چکیده انگلیسی
After its investigation of cross-contamination from arthroscopic shavers, the US Food and Drug Administration issued an alert to hospitals about medical device reprocessing methods. In response to this, a team of risk management and instrument room personnel at a university hospital undertook a project that tested the manufacturer's recommended cleaning methods for surgical instruments with the objective of determining the efficacy of automated instrument reprocessing and identifying a process that would produce a verifiably clean instrument after the cleaning process is performed. The quality improvement project focused on suction tips because they are used in most surgical procedures, are exposed to high levels of organic debris, and are difficult to clean. A variety of suction tips were cleaned and tested with a variety of processes and products to determine best instrument cleaning practices. Results of the project were eye-opening-debris was found where debris should not be, and the manufacturer's recommended cleaning methods-the current practices-were not effective.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: AORN Journal - Volume 96, Issue 2, August 2012, Pages 152-162
نویسندگان
, , , , , , , , ,