کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2643926 1137885 2006 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Electrosurgical burns in pediatric patients undergoing liver resection with saline-enhanced radiofrequency technology
موضوعات مرتبط
علوم پزشکی و سلامت پرستاری و مشاغل بهداشتی پرستاری
پیش نمایش صفحه اول مقاله
Electrosurgical burns in pediatric patients undergoing liver resection with saline-enhanced radiofrequency technology
چکیده انگلیسی
ABSTRACT
- USE OF SALINE-ENHANCED radiofrequency (RF) technology reduces blood loss in infants undergoing liver resection. Radiofrequency systems continuously deliver much higher currents for longer periods of time than conventional coagulation instruments and thus increase the risk of complications.
- THREE INFANTS undergoing liver resection with the newer RF technology at one facility experienced electrosurgical unit (ESU) dispersive pad burns. As a result, an experiment was performed on animal subjects to determine best ESU dispersive pad placement on infants and to gather data on device impedance.
- RECOMMENDATIONS to prevent burns include central placement of the ESU dispersive pad and activating warming devices only after resection is complete. AORN J 83 (March 2006) 658-664.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: AORN Journal - Volume 83, Issue 3, March 2006, Pages 657-660, 662-664