کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2645295 1138489 2011 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Comparison of oxygen saturation values and measurement times by pulse oximetry in various parts of the body
موضوعات مرتبط
علوم پزشکی و سلامت پرستاری و مشاغل بهداشتی پرستاری
پیش نمایش صفحه اول مقاله
Comparison of oxygen saturation values and measurement times by pulse oximetry in various parts of the body
چکیده انگلیسی

The aim of this study, which included 40 patients, was to compare the values pulse oximetry and the measurement times in various regions of the body. Data were analyzed using intraclass correlation coefficient test and paired-sample test. The confidence power value was found to be .81 for the comparison of oxygen saturation values by arterial blood gas analysis and measurement by the forehead probe. It was found that the time for oxygen saturation measurement using the forehead probe was shorter than those using the finger and toe probes.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Nursing Research - Volume 24, Issue 4, November 2011, Pages e39–e43