کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2836946 1164872 2015 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Primary endpoint results of the OMEGA Study: One-year clinical outcomes after implantation of a novel platinum chromium bare metal stent ★
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی پزشکی مولکولی
پیش نمایش صفحه اول مقاله
Primary endpoint results of the OMEGA Study: One-year clinical outcomes after implantation of a novel platinum chromium bare metal stent ★
چکیده انگلیسی


• The OMEGA study evaluated a novel platinum chromium bare metal stent.
• OMEGA enrolled 328 patients at 37 sites (US and Europe).
• The primary endpoint of 9 month target lesion failure was 11.5%.
• One-year event rates were low including an ST rate of 0.6% at 12 months.

Background/purposeBare metal stents (BMS) have similar rates of death and myocardial infarction (MI) compared to drug-eluting stents (DES). DES lower repeat revascularization rates compared to BMS, but may have higher rates of late stent thrombosis (ST) potentially due to impaired endothelialization requiring longer dual anti-platelet therapy (DAPT). OMEGA evaluated a novel BMS designed to have improved deliverability and radiopacity, in comparison to currently available platforms.Methods/materialsOMEGA was a prospective, multicenter, single-arm study enrolling 328 patients at 37 sites (US and Europe). Patients received the OMEGA stent (bare platinum chromium element stent) for the treatment of de novo native coronary artery lesions (≤ 28 mm long; diameter ≥ 2.25 mm to ≤ 4.50 mm). The primary endpoint was 9-month target lesion failure (TLF: cardiac death, target vessel-related MI, target lesion revascularization [TLR]) compared to a prespecified performance goal (PG) based on prior generation BMS. All major cardiac events were independently adjudicated. DAPT was required for a minimum of 1 month post procedure.ResultsIn the OMEGA study, the mean age was 65; 17% had diabetes mellitus. The primary endpoint was met; 9 month TLF rate was 11.5%, and the upper 1-sided 95% confidence bound of 14.79% was less than the prespecified PG of 21.2% (p < 0.0001). One-year event rates were low including a TLF rate of 12.8% and an ST rate of 0.6% at 12 months.ConclusionsOne-year outcomes of OMEGA show low rates of TLF, revascularization and ST. This supports safety and efficacy of the OMEGA BMS for the treatment of coronary artery disease.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Cardiovascular Revascularization Medicine - Volume 16, Issue 2, March 2015, Pages 65–69
نویسندگان
, , , , , , , , ,