کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2848100 1167405 2008 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Bench study of auto-CPAP devices using a collapsible upper airway model with upstream resistance
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی فیزیولوژی
پیش نمایش صفحه اول مقاله
Bench study of auto-CPAP devices using a collapsible upper airway model with upstream resistance
چکیده انگلیسی
The aim of this study was to investigate the response of auto-CPAP devices to respiratory events (apnea, hypopnea, flow-limitation and snoring) on the same condition using a physiological upper airway model. The hypothesis of this study is that collapsibility of the flow-limiting collapsible segment of the airway is influenced by the upstream airway resistance. Five auto-CPAP devices, AutoSet®T, AutoSet® Spirit™, Goodnight®420E, PV10i and REMstar®Auto were evaluated. Apnea: all the devices increased the auto-CPAP level, while AutoSet®T and AutoSet® Spirit™ did not respond to apnea for 30 s. Hypopnea: all the devices except the AutoSet®T and Goodnight®420E increased pressure. Flow-limitation: all the devices except the PV10i and REMstar®Auto increased pressure. Snoring: the snoring sounds disappeared when REMstar®Auto and PV10i were used, and the Goodnight®420E lowered the level of snoring. In conclusion, the response of auto-CPAP devices to respiratory events differed. Collapsible upper airway model with upstream resistance is useful for the first-step assessment of auto-CPAP devices.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Respiratory Physiology & Neurobiology - Volume 162, Issue 1, 30 June 2008, Pages 48-54