کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
295652 511564 2007 12 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Experimental–computational solution to fatigue induced microcrack source localization in cemented hip arthroplasty models
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی عمران و سازه
پیش نمایش صفحه اول مقاله
Experimental–computational solution to fatigue induced microcrack source localization in cemented hip arthroplasty models
چکیده انگلیسی

The present work examined the effects of signal velocity models, system equation damping, number of sensors, and system singularity on experimental–computational solutions to locate fatigue-induced microcracks in cemented total hip arthroplasty (THA). The following results were found. (1) Variable velocity model is a better approach for localizing microcracks in THA constructs. (2) The optimal range of system damping is between 0.001 and 0.002. (3) The rounding error of the final solution can be limited to the level of 10−2 when the numerical precision is taken to six decimal places for each element in the system matrix. In addition, proper use of system damping can reduce system singularity.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: NDT & E International - Volume 40, Issue 5, July 2007, Pages 378–389