Accurate measurements of dissolved O2 as a function of time have numerous chemical and biological applications. The Pd (II) complex of meso-tetra-(4-sulfonatophenyl)-tetrabenzoporphyrin (Pd phosphor) was used for this purpose. Detection is based on the principle that the phosphorescence of this oxygen probe is inversely related to dissolved O2 (O2 quenches the phosphorescence). Biologic samples containing the Pd phosphor were flashed (10/s) with a peak output of 625 nm; emitted light was detected at 800 nm. Amplified pulses of phosphorescence were digitized at 1–2 MHz using an analog/digital converter (PCI-DAS 4020/12 I/O Board) with outputs ranging from 1 to 20 MHz. Assessment revealed a customized program was necessary. Pulses were captured using a developed software at 0.1–4 MHz, depending on the speed of the computer. O2 concentration was calculated by fitting to an exponential the decay of the phosphorescence. Twelve tasks were identified, which allowed full control and customization of the data acquisition, storage and analysis. The program used Microsoft Visual Basic 6 (VB6), Microsoft Access Database 2007, and a Universal Library component that allowed direct reading from the PCI-DAS 4020/12 I/O Board. It involved a relational database design to store experiments, pulses and pulse metadata, including phosphorescence decay rates. The method permitted reliable measurements of cellular O2 consumption over several hours.
Journal: Computer Methods and Programs in Biomedicine - Volume 100, Issue 3, December 2010, Pages 265–268