Application of multiobjective neural predictive control to biventricular assistance using dual rotary blood pumps
ترجمه فارسی عنوان
استفاده از کنترل پیش بینی چندضلعی عصبی به کمک دو طرفه با استفاده از پمپ های دوار دوتایی
موضوعات مرتبط
مهندسی و علوم پایه مهندسی کامپیوتر پردازش سیگنال
چکیده انگلیسی


- A multiobjective neural network predictive controller is proposed for a BiVAD.
- The controller is able to alleviate pulmonary congestion and ventricular suction.
- The proposed controller outperformed Dual-Independent FS and constant speed modes.
- The proposed controller remained robust in the presence of sensor noise and delays.

Rotary blood pumps are used to provide mechanical circulatory support to the failing heart in patients who are ineligible or waiting for a transplant. One of the major challenges when implementing two rotary blood pumps for biventricular support is the difficulty in maintaining pulmonary and systemic circulatory volume balance. In this study, a novel multiobjective neural predictive controller (MONPC) hybridized with a preload-based Frank-Starling-like controller (PFS) has been proposed for a dual rotary blood pump biventricular assist device in two different configurations: PFSL-MONPCR and MONPCL-PFSR. The flow rate of one pump is regulated by PFS as a function of preload, while the other pump is controlled by MONPC, which is intended to meet cardiac demand, avoid pulmonary congestion and ventricular suction. A comparative assessment was performed between the proposed controllers and a Dual Independent Frank-Starling-like control system (DI-FS) as well as a constant speed controller. The numerical simulation results showed that MONPCL-PFSR helped unload the congested left ventricle while maintaining high cardiac output during exercise. In contrast, improper flow regulation by DI-FS has resulted in pulmonary congestion. During blood loss, PFSL-MONPCR delivered the lowest suction risk, as compared to the constant speed mode, which produced negative right ventricular preload. When sensor noise and time delays were introduced in the flow and end-diastolic pressure signals, the proposed controllers were able to respond with adequate robustness during the transition from rest to exercise. This study demonstrated that the proposed controllers are superior in matching the pump flow with the cardiac demand without causing hemodynamic instabilities.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biomedical Signal Processing and Control - Volume 39, January 2018, Pages 81-93
نویسندگان
, , , , , , , ,