کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4992375 1457263 2017 10 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
On the pulsatile flow through a coronary bifurcation with stent
ترجمه فارسی عنوان
در جریان خونریزی از طریق بیرون زدن کرونر با استنت
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی جریان سیال و فرایندهای انتقال
چکیده انگلیسی

The present study investigates pulsatile flow through models of junctions corresponding to geometries most usually observed in the major coronary artery, with and without stent. A transparent experimental device is used to obtain PIV measurements of the instantaneous flow fields with respect to both meridian and cross sections. They clearly show the persistence of a strong secondary flow, even when a stent is implanted. A numerical analysis is then performed from the experimental data to deduce the circulation of the cross velocity field. The balance of the circulation within the three measurement sections can be interpreted as a factor measuring the Wall Shear Stress (WSS). It is a relatively simple way to estimate the total WSS caused by changing flow due to the presence of a stent.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: European Journal of Mechanics - B/Fluids - Volume 61, Part 1, January–February 2017, Pages 177-186
نویسندگان
, , , ,