کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
503945 864251 2016 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Phantom study of an ultrasound guidance system for transcatheter aortic valve implantation
موضوعات مرتبط
مهندسی و علوم پایه مهندسی کامپیوتر نرم افزارهای علوم کامپیوتر
پیش نمایش صفحه اول مقاله
Phantom study of an ultrasound guidance system for transcatheter aortic valve implantation
چکیده انگلیسی


• A guidance system for TAVI using ultrasound and magnetic tacking was developed.
• Eliminates the use of nephrotoxic contrast and ionizing radiation.
• In phantom experiment, similar accuracy in deployment depth as with fluoroscopy.
• A new delivery tool with integrated ICE further improves deployment accuracy.

A guidance system using transesophageal echocardiography and magnetic tracking is presented which avoids the use of nephrotoxic contrast agents and ionizing radiation required for traditional fluoroscopically guided procedures. The aortic valve is identified in tracked biplane transesophageal echocardiography and used to guide stent deployment in a mixed reality environment. Additionally, a transapical delivery tool with intracardiac echocardiography capable of monitoring stent deployment was created. This system resulted in a deployment depth error of 3.4 mm in a phantom. This was further improved to 2.3 mm with the custom-made delivery tool. In comparison, the variability in deployment depth for traditional fluoroscopic guidance was estimated at 3.4 mm.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Computerized Medical Imaging and Graphics - Volume 50, June 2016, Pages 24–30
نویسندگان
, , , , , , ,