Non-invasive beat-to-beat cardiac output monitoring by an improved method of transthoracic bioimpedance measurement
چکیده انگلیسی

The report describes a method of impedance cardiography using an improved estimate of thoracic volume. The formulas and their implementation in hardware and software are explained and new shortband electrodes are described which generate a good homogeneous thoracic field. Examples of stroke volume and cardiac output curves underline the capabilities of the monitoring system “Task Force®Force® Monitor”. In several experiments, results are compared to thermodilution as well as to BioZ measurements: the new method excels in comparison with thermodilution and is comparable to the BioZ device. Compared to traditional electrodes, the new shortband electrodes are shown to provide better reproducibility.