کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5431172 1397440 2006 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Comparison of the quantitative results corrected by fundamental parameter method and difference calibration specimens in X-ray fluorescence spectrometry
موضوعات مرتبط
مهندسی و علوم پایه شیمی طیف سنجی
پیش نمایش صفحه اول مقاله
Comparison of the quantitative results corrected by fundamental parameter method and difference calibration specimens in X-ray fluorescence spectrometry
چکیده انگلیسی

Three kinds of standard specimens, including low-alloy steels (Japanese iron and steel CRM), glass from NIST and Chinese geological reference materials were analyzed by wavelength dispersive X-ray fluorescence spectrometry as unknown samples. The fundamental parameter method was used to correct the matrix effect and different specimens including pure element specimens and type specimens were employed as calibration samples. The quantitative results were compared in order to study the influence of calibration samples on the accuracy of quantitative results. The study showed that the fundamental parameter method could correct the matrix effects very well and relatively accurate quantitative results could be obtained, corrected by simple specimens, in case of type calibration specimens' absence.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Quantitative Spectroscopy and Radiative Transfer - Volume 97, Issue 1, January 2006, Pages 68-74
نویسندگان
, , , , ,