کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
549600 872389 2010 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A portable device for the assessment of dynamic visual acuity
موضوعات مرتبط
مهندسی و علوم پایه مهندسی کامپیوتر تعامل انسان و کامپیوتر
پیش نمایش صفحه اول مقاله
A portable device for the assessment of dynamic visual acuity
چکیده انگلیسی

Dynamic visual acuity (DVA) thresholds are among the few visual functions predictive of automobile crashes. DVA is also sensitive to alcohol and aging. However, measuring DVA is awkward because there is no standardized, efficient, flexible apparatus for DVA assessment. In this project, we developed a prototype of an automated, portable DVA system using a low-energy laser, and we compared this laser DVA with the traditional device in two within-subjects, repeated measures designs. The two studies included 48 participants (22 males and 26 females with an average age of 18.33 years). The most important findings were that: (1) retest reliabilities of the two DVA devices were comparable and higher with the laser; (2) average correlations between the two devices were r = 0.62 (p < 0.01) and r = 0.65 (p < 0.01) for the two designs respectively; and (3) after correction for reliability attenuation these improved to r = 0.92 and r = 0.78. These findings indicate that a flexible DVA laser device can be developed to measure the same construct as the more traditional bulky DVA device.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Ergonomics - Volume 41, Issue 2, March 2010, Pages 266–273
نویسندگان
, ,