Experimental investigation on minimum frame rate requirements of high-speed videoendoscopy for clinical voice assessment
ترجمه فارسی عنوان
بررسی تجربی در مورد حداقل الزامات نرخ فریم در ویدئوهای آندوسکوپی با سرعت بالا برای ارزیابی صدای بالینی
کلمات کلیدی
ویدئودوسکوپی با سرعت بالا، نرخ فریم، ارزیابی صدای بالینی، تصویربرداری حنجره
موضوعات مرتبط
مهندسی و علوم پایه مهندسی کامپیوتر پردازش سیگنال
چکیده انگلیسی


• Voice assessment was studied using laryngeal imaging via high-speed videoendoscopy
• Frame rates impact the assessment of clinical vocal fold vibratory features
• Glottal edge, mucosal wave, aperiodicity and contact are most sensitive to frame rate
• The recommended frame rates for laryngeal imaging should be 8000 frames per second
• The minimum frame rates for laryngeal imaging should be 4000 frames per second

This study investigated the impact of high-speed videoendoscopy (HSV) frame rates on the assessment of nine clinically relevant vocal-fold vibratory features. Fourteen adult patients with voice disorder and 14 adult normal controls were recorded using monochromatic rigid HSV at a rate of 16,000 frames per second (fps) and spatial resolution of 639 × 639 pixels. The 16,000-fps data were downsampled to 16 other rate denominations. Using paired comparisons design, nine common clinical vibratory features were visually compared between the downsampled and the original images. Three raters reported the thresholds at which: (1) a detectable difference between the two videos was first noticed, and (2) differences between the two videos would result in a change of clinical rating. Results indicated that glottal edge, mucosal wave magnitude and extent, aperiodicity, contact and loss of contact of the vocal folds were the vibratory features most sensitive to frame rate. Of these vibratory features, the glottal edge was selected for further analysis, due to its higher rating reliability, universal prevalence and consistent definition. Rates of 8000 fps were found to be free from visually perceivable feature degradation, and for rates of 5333 fps, degradation was minimal. For rates of 4000 fps and higher, clinical assessments of glottal edge were not affected. Rates of 2000 fps changed the clinical ratings in over 16% of the samples, which could lead to inaccurate functional assessment.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biomedical Signal Processing and Control - Volume 17, March 2015, Pages 21–28