کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
607694 1454592 2013 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Response of a microcapillary impedimetric transducer to changes in surface conductance at liquid/solid interface
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی شیمی کلوئیدی و سطحی
پیش نمایش صفحه اول مقاله
Response of a microcapillary impedimetric transducer to changes in surface conductance at liquid/solid interface
چکیده انگلیسی


• A new microcapillary impedimetric sensor for surface charge monitoring is proposed.
• The mechanism is based on changes in conductivity due to changes in surface charge.
• The device is useful for real time adsorption kinetics monitoring.

A new device based on an interdigitated electrode array with electrode digits located at the bottom of microcapillaries is presented. Microcapillaries formed in silicon dioxide are 3 μm wide, 4 μm high and are open at the top, so that in contact with an electrolyte solution the AC current flows from one capillary to another and is significantly affected by changes in surface conductivity at the SiO2/electrolyte interface. The effect of charged polyelectrolyte layers electrostatically assembled on the sensor surface on the surface conductivity in solutions with different KCl concentration is presented. From measured impedance spectra polyelectrolyte adsorption curve is determined. The device is shown to be useful for real time adsorption kinetics monitoring.

Figure optionsDownload high-quality image (101 K)Download as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Colloid and Interface Science - Volume 403, 1 August 2013, Pages 151–156
نویسندگان
, ,