Wear comparison between a dual mobility total hip prosthesis and a typical modular design using a hip joint simulator
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
مهندسی شیمی
شیمی کلوئیدی و سطحی
پیش نمایش صفحه اول مقاله
چکیده انگلیسی
The wear behaviour of a new dual mobility total hip design was compared with that of a modular design using the 12-station anatomic hip joint simulator HUT-4. In addition, two positions of the acetabular shells were compared, at 45° and 60° abduction. The acetabular insert material was conventional ultra-high molecular weight polyethylene (UHMWPE) in both designs, and the femoral head material was stainless steel. The differences in the mean wear rates between the two designs in either position, and between the two positions in either design were not statistically significant. The wear rates were of the order of 20 mg per one million cycles.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Wear - Volume 268, Issues 3â4, 4 February 2010, Pages 617-621
Journal: Wear - Volume 268, Issues 3â4, 4 February 2010, Pages 617-621
نویسندگان
Vesa Saikko, Ming Shen,