کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
743254 | 894348 | 2009 | 5 صفحه PDF | دانلود رایگان |

A disposable electrochemical immunofiltration test strip for the rapid detection of α-fetoprotein (AFP) was developed. The test strip was constructed by assembly of screen-printed carbon electrodes, absorption-water pad, nitrocellulose membrane modified by anti-AFP antibody and glass fiber membrane conjugated with ferrocene monocarboxylic acid (FC) labeling AFP. The analytical system utilizes flow-through immunofiltration and competitive immunoassay techniques in combination with an amperometric sensor. The parameters affecting the immunoassay such as selection of filter membrane, membrane pore-size, and antibody binding capacity were investigated and optimized. The immunofiltration system allows us to specifically and directly detect AFP in serum with a low detection limit of 6 ng/mL. The working range is from 6 to 500 ng/mL with an overall analysis time of 5 min for one sample. This electrochemical immunoassay system enabled us to construct a novel point-of-care testing device for the monitoring of biomarker including AFP.
Journal: Sensors and Actuators B: Chemical - Volume 141, Issue 2, 7 September 2009, Pages 436–440