کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
743660 894365 2008 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Detection of Escherichia coli using CMOS array photo sensor-based enzyme biochip detection system
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Detection of Escherichia coli using CMOS array photo sensor-based enzyme biochip detection system
چکیده انگلیسی

This work presents optical enzyme detection system based on the CMOS array photo sensor and 1 × 3 polymeric enzyme biochip for detecting Escherichia coli in a one-step procedure. This assay, using 4-methylumbelliferyl-β-d-glucuronide (MUG) as a fluorogenic substrate, had a detection limit of 0.1 U/ml for β-glucuronidase (GUD), which was approximately equal to a cell concentration of 106 CFU/ml of E. coli. MUG was incorporated into lauryl tryptose broth at a final concentration of 100 μg/ml for immediate verification of the presence of E. coli in 1 × 3 polymeric enzyme biochip. The 40 strains of E. coli studied all produced GUD. Of another 36 strains of bacteria tested, one strain (Salmonella choleraesuis subsp. choleraesuis) yielded very small amounts of GUD after 24 h incubation. The optical enzyme detection system was sensitive and rapid.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Sensors and Actuators B: Chemical - Volume 133, Issue 2, 12 August 2008, Pages 561–564
نویسندگان
,