کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
755120 | 895926 | 2009 | 9 صفحه PDF | دانلود رایگان |

This paper presents the theory, design, and validation of a microphone calibrator used to simultaneously calibrate the amplitudes of multiple microphones on a single probe. The probe uses four 6 mm diameter electret microphones to acquire the data needed to compute acoustic energy density. This probe has prompted the need for simultaneous, multi-microphone amplitude calibration. The calibration process simultaneously subject each microphone on the probe to the same known acoustic pressure using four equal-length, small-diameter tubes connected to a single excitation source. A reference microphone connected to a fifth tube is used to calibrate the microphones. Test results show that the calibrator can calibrate each probe microphone within ±0.5 dB up to 2 kHz, and within ±1 dB up to 4.9 Hz with a confidence level of 95%.
Journal: Applied Acoustics - Volume 70, Issue 6, June 2009, Pages 790–798