Determination of As, Sb and Hg in water samples by flow injection coupled HR CS ETAAS with an in situ hydride generator
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Determination of As, Sb and Hg in water samples by flow injection coupled HR CS ETAAS with an in situ hydride generator
چکیده انگلیسی
This method has proven to have a high sensitivity with detection limits of 0.25 μg L−1, 0.003 μg L−1 and 0.22 μg L−1 for As, Sb and Hg, respectively. Besides, the proposed method was applied in the sequential determination of the 3 elements mentioned above with a sample throughput of about 16.7 h−1, thereby, reducing the time of analysis and the volume of reagents and sample required. This method was applied to determine the target elements in a TMDA 54.5 and TM 24.3 certified reference materials, and seawater collected from Malaga Bay. The results showed good agreement with the certified values.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Microchemical Journal - Volume 138, May 2018, Pages 109-115