کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8130626 1523213 2015 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Focused intravascular ultrasonic probe using dimpled transducer elements
ترجمه فارسی عنوان
متمرکز پروب اولتراسونیک داخل عضلانی با استفاده از عناصر مبدل دمیده شده
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم آکوستیک و فرا صوت
چکیده انگلیسی
High-frequency focused intravascular ultrasonic probes were fabricated in this study using dimple technique based on PMN-PT single crystal and lead-free KNN-KBT-Mn ceramic. The center frequency, bandwidth, and insertion loss of the PMN-PT transducer were 34 MHz, 75%, and 22.9 dB, respectively. For the lead-free probe, the center frequency, bandwidth, and insertion loss were found to be 40 MHz, 72%, and 28.8 dB, respectively. The ultrasonic images of wire phantom and vessels with good resolution were obtained to evaluate the transducer performance. The −6 dB axial and lateral resolutions of the PMN-PT probe were determined to be 58 μm and 131 μm, respectively. For the lead-free probe, the axial and lateral resolutions were found to be 44 μm and 125 μm, respectively. These results suggest that the mechanical dimpling technique has good potential in preparing focused transducers for intravascular ultrasound applications.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ultrasonics - Volume 56, February 2015, Pages 227-231
نویسندگان
, , , , , , , , ,