Correction of quenching effect of a small size OSL dosimeter using Eu:BaFBr and Ce:CaF2
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Correction of quenching effect of a small size OSL dosimeter using Eu:BaFBr and Ce:CaF2
چکیده انگلیسی
To accurately estimate an actual dose during radiotherapy treatment, dosimeters are required to be inserted into the affected region in a patient's body. Therefore, we are developing a small size dosimeter consisting of optical fibers and optically stimulated luminescence (OSL) elements. We fabricated two types of small size dosimeters using Eu:BaFBr and Ce:CaF2. We measured the Bragg peak of high energy carbon ions in a water equivalent material. OSL materials showed the quenching effect when irradiated by high linear energy transfer (LET) particles. The sensitivity of Eu:BaFBr and Ce:CaF2 monotonically decreased with different coefficients. The ratio of the signal intensity of these two OSL materials can uniquely determine the sensitivity of each OSL material. We corrected the quenching effect of the OSL dosimeters in the spread out Bragg peak with this relationship.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiation Measurements - Volume 106, November 2017, Pages 246-251