Deep brain stimulation failure due to external cardioversion in a patient with Parkinson's disease
ترجمه فارسی عنوان
نارسایی تحریک مغزی عمیق به علت کاردیوورژی خارجی در بیمار مبتلا به بیماری پارکینسون
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی تحقیقات سرطان
چکیده انگلیسی
We report a case of deep brain stimulation (DBS) hardware failure due to emergently performed subcutaneous coronary angioplasties complicated by cardioversion for rapid worsening of angina pectoris and some trouble shooting problems emerged after invasive cardiovascular procedures. The patient with prior implantation of permanent pacemaker due to vasovagal syndrome underwent successful left-sided unilateral electrode implantation into the subthalamic nucleus. During 21 months follow-up period the patient experienced 2 times episodes of aggravation of unstable angina pectoris 15 and 21 months respectively, which necessities emergent coronary angioplasties. After the first emergently performed coronary angioplasty with cardioversion the interrogation of DBS system revealed the depletion of an internal pulse generator (IPG). The secondly performed coronary angioplasty complicated by ventricular tachyarrhythmia with DBS system switched on during emergent cardioversion resulted in partial dysfunction of DBS electrode. Patients harboring cardiovascular implantable electronic devices (CIEDs) and DBS systems require special attention and good cooperation of neurosurgeons, interventional cardiologist, and neurologist. Some emergently performed invasive cardiovascular procedures which necessities cardioversion may cause DBS hardware failure with subsequent worsening of movement disorder symptoms.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Neurologia i Neurochirurgia Polska - Volume 51, Issue 4, July–August 2017, Pages 324-330
نویسندگان
, , , ,