کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10051236 1598561 2005 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Measurement of Whole Blood Tacrolimus Level by High Performance Liquid Chromatography Tandem Mass Spectrometry in Renal Transplant Recipients - A Single Center Perspective
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیماری‌های کلیوی
پیش نمایش صفحه اول مقاله
Measurement of Whole Blood Tacrolimus Level by High Performance Liquid Chromatography Tandem Mass Spectrometry in Renal Transplant Recipients - A Single Center Perspective
چکیده انگلیسی
For tacrolimus concentrations within the recommended therapeutic range, the concentration measured by HPLC-MS/MS was statistically significantly lower than that measured by MEIA, but the difference was not clinically significant. Introduction of the more specific HPLC-MS/MS method does not require adjustment in the recommended tacrolimus trough concentration or the AUC12 level estimated by our abbreviated regression equation.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Hong Kong Journal of Nephrology - Volume 7, Issue 2, October 2005, Pages 65-69
نویسندگان
, , , , , , , ,