کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10407423 892968 2013 13 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Risø calibration quartz - A challenge for β-source calibration. An applied study with relevance for luminescence dating
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی کنترل و سیستم های مهندسی
پیش نمایش صفحه اول مقاله
Risø calibration quartz - A challenge for β-source calibration. An applied study with relevance for luminescence dating
چکیده انگلیسی
Risø calibration quartz (RCQ) is used worldwide in luminescence laboratories to calibrate β-sources. RCQ is advantageous because the inherent γ-dose is determined conveniently with a blue light stimulated luminescence (BLSL) single aliquot regeneration (SAR) protocol and dose equivalents (DEs) are highly reproducible with standard errors of the mean of usually <2%. We report on a series of tests on the recoverability of known β-doses from RCQ and on DE-determination of γ-dosed RCQ with varying SAR-measurement parameters. Evaluated alone, these raise doubt whether the precise calibration results are systematically correct. However, it is likely that cutheat and preheat adjustment of 'quick calibration' as applied by staff members of the Radiation Research Division of the Technical University of Denmark when installing a new luminescence reader in a customer's laboratory are appropriate to minimize systematic errors. This finding is suggested by comparative calibration of a β-source with RCQ and another γ-irradiated quartz.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Measurement - Volume 46, Issue 7, August 2013, Pages 2238-2250
نویسندگان
, ,