Cell-mediated retraction versus hemodynamic loading - A delicate balance in tissue-engineered heart valves
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
Cell-mediated retraction versus hemodynamic loading - A delicate balance in tissue-engineered heart valves
چکیده انگلیسی
Stress generation by cells was measured using an earlier described in vitro model system, mimicking the culture process of TEHVs. The stress imposed by the blood pressure during diastole on a valve leaflet was determined using finite element modeling. Results show that for both pulmonary and systemic pressure, the stress imposed on the TEHV leaflets is comparable to the stress generated in the leaflets. As the stresses are of similar magnitude, it is likely that the imposed stress cannot counteract the generated stress, in particular when taking into account that hemodynamic loading is only imposed during diastole. This study provides a rational explanation for the retraction found in preclinical studies of TEHVs and represents an important step towards understanding the retraction process seen in TEHVs by a combined experimental and computational approach.