کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10432261 910242 2014 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Passive aortic counterpulsation: Biomechanical rationale and bench validation
ترجمه فارسی عنوان
انفجار آئورت منفی: اعتبار سنجی بیومکانیک و پایه
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
چکیده انگلیسی
Using a cardiovascular simulator and changing the reservoir pressure, a systolic not negligible (7.8 ml) gas volume exchange between the intra-aortic balloon and the reservoir was measured. The same cardiovascular simulator was used to demonstrate an increase in stroke volume in three conditions of progressive ventricular failure, by detecting a change in systolic and diastolic arterial pressures and stroke volume (SV). The maximal arterial pressure always decreased and the diastolic pressure increased. The SV increased up to 7.8%, demonstrating an arterial elastance reduction and better ventricular-aortic mechanical matching and energy transfer.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Biomechanics - Volume 47, Issue 7, 7 May 2014, Pages 1618-1625
نویسندگان
, , , , ,