کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10434515 910418 2005 11 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Effect of segmentation errors on 3D-to-2D registration of implant models in X-ray images
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
Effect of segmentation errors on 3D-to-2D registration of implant models in X-ray images
چکیده انگلیسی
Although the silhouette of a metallic object is often clearly visible in an X-ray image, adjacent tissue and occlusions can make the exact location of the silhouette contour difficult to determine in places. Occlusion can occur when another object (such as another implant component) partially blocks the view of the object of interest. In this paper, we argue that common methods for segmentation can produce errors in the location of the 2D contour, and hence errors in the resulting 3D estimate of the pose. We show, on a typical fluoroscopy image of a knee implant component, that interactive and automatic methods for segmentation result in segmented contours that vary significantly. We show how the variability in the 2D contours (quantified by two different metrics) corresponds to variability in the 3D poses. Finally, we illustrate how traditional segmentation methods can fail completely in the (not uncommon) cases of images with occlusion.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Biomechanics - Volume 38, Issue 2, February 2005, Pages 229-239
نویسندگان
, , , ,