کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10532866 961756 2010 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Kinetic analysis and fragment screening with Fujifilm AP-3000
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Kinetic analysis and fragment screening with Fujifilm AP-3000
چکیده انگلیسی
We evaluated the performance of Fujifilm's new AP-3000 surface plasmon resonance biosensor for kinetic analysis and fragment screening. Using carbonic anhydrase II as a model system, we characterized a set of 10 sulfonamide-based inhibitors that range in molecular mass from 98 to 341 Da and approximately 10,000-fold in affinity (0.4 mM to 20 nM). Although the data collected from the AP-3000 were generally similar to those collected using a Biacore T100, the AP-3000's stop-flow analyte delivery system complicated the shapes of the association- and dissociation-phase binding responses. We illustrate how reasonable estimates of the kinetic rate constants can be extracted from AP-3000 data by limiting data analysis to only the regions of the responses collected during flow conditions. We also provide an example of the results obtained for a fragment-screening study with the AP-3000, which is the ideal application of this technology.