کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10706093 1023307 2005 11 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Reversible ageing effects in cryogenically cooled infrared filter radiometers
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم فیزیک اتمی و مولکولی و اپتیک
پیش نمایش صفحه اول مقاله
Reversible ageing effects in cryogenically cooled infrared filter radiometers
چکیده انگلیسی
In this paper we report the observation of drifts in the responsivity of cryogenically cooled InSb detector-based infrared filter radiometers which have very strong wavelength dependence. These drifts can result in the increase or decrease of the response of the filter radiometers by over 5%. The origin of these variations was investigated and was shown to arise due to a thin film of ice formed on the multi-layer bandpass filter used to define the spectral response of the filter radiometer. The thin layer of ice interacts with the characteristics of the filter (which itself consists of a number of thin layers) and modifies the filter spectral transmission thus modifying the response of the filter radiometer of which the filter is part of. These observations are particularly relevant to space instruments which use infrared filter radiometers for earth observation. Debris from the spacecraft engines is known to accumulate on cold surfaces of instruments carried on board. The deposition of this debris on cold filters can modify the spectral response of the instruments, which use these filters to define a spectral response.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Infrared Physics & Technology - Volume 46, Issue 4, April 2005, Pages 339-349
نویسندگان
, , ,