Comparison of two devices for the treatment of keloid scars with the use of intralesional cryotherapy: An experimental study
چکیده انگلیسی
In conclusion, the argon gas device displayed a lower end temperature and a faster freezing rate in vivo compared to the liquid nitrogen device. Although this resulted in lower recurrence rates for the argon gas device, more hypopigmentation was seen compared to the liquid nitrogen device following treatment. Finally, the low outer surface temperatures measured with both devices, suggest that some hypopigmentation following treatment is inevitable.