Implementation of the on-chip immunoassay for α-fetoprotein (AFP)–L3% was achieved using a fully automated microfluidic instrument platform that will prepare the chip and run the assay with a total assay time of less than 10 min. Reagent/sample mixing, concentration, and reaction in microfluidic channels occur by the electrokinetic analyte transport assay (EATA) technique, enabling the integration of all assay steps on-chip. The determination of AFP–L3%, a biomarker for hepatocellular carcinoma, was achieved by the presence of Lens culinaris agglutinin in the separation channel, causing separation of the fucosylated isoform, AFP–L3, from the nonfucosylated AFP–L1 by lectin affinity electrophoresis. Laser-induced-fluorescence (LIF) detection was used to quantitate the labeled immunocomplexes. The limit of detection (LOD) was 0.1 ng/ml AFP, and assay precision of less than 2% coefficient of variation (CV) was obtained for quantitation from 24 to 922 ng/ml total AFP in spiked serum samples. Assay precision of less than 3% CV was obtained for AFP–L3% measurements from 8.5 to 81%. Furthermore, good correlation of test results for 68 patient serum samples with a commercially available reference method (LiBASys assay for AFP–L3%) was obtained, with r2 = 0.981 and slope = 1.03.
Journal: Analytical Biochemistry - Volume 388, Issue 2, 15 May 2009, Pages 306–311