Correction of multi-collector-ICP-MS instrumental biases in high-precision uranium–thorium chronology
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Correction of multi-collector-ICP-MS instrumental biases in high-precision uranium–thorium chronology
چکیده انگلیسی

We present a method for the calibration of in-house thorium isotope standards for use in U–Th chronology, along with an assessment of instrumental biases on the Nu Plasma multi-collector inductively-coupled-plasma mass spectrometer. Uranium and thorium mass fractionation is found to be very strongly coupled, with no evidence of differences in mass bias behaviour. Both Th-isotope and U–Th fractionation are most closely approximated by a linear-law fractionation model. Any deviations from linear law are at the sub-permil level. We also investigate potential biases introduced by intensity-dependant, and ion-beam-path dependant variations in ion-counter response. At count rates <∼350,000 counts per second, intensity-dependant gain variation can be accounted for entirely by dead-time, while beam-path differences can cause gain variation of up to ∼0.8% for different Th isotopes. If this beam-path bias is to be adequately corrected, ion-counter measurements of Th-isotope ratios must be made with reference to bracketing Th-isotope standards measured in the same configuration.

New in-house thorium standards have been calibrated and used to assess instrumental biases in thorium isotope measurement on a Nu Plasma ICP-MS.Figure optionsDownload high-quality image (213 K)Download as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: International Journal of Mass Spectrometry - Volume 295, Issues 1–2, 15 July 2010, Pages 26–35
نویسندگان
, ,