کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1228058 968445 2009 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A multi-syringe flow injection system for the spectrophotometric determination of trace levels of iron in waters using a liquid waveguide capillary cell and different chelating resins and reaction chemistries
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A multi-syringe flow injection system for the spectrophotometric determination of trace levels of iron in waters using a liquid waveguide capillary cell and different chelating resins and reaction chemistries
چکیده انگلیسی

A method integrating a long waveguide capillary cell with a preconcentration resin in a multi-syringe flow injection analysis (MSFIA) system for iron determination in waters was developed. The determination of iron is based on a colorimetric reaction and two reagents were tested, ferrozine and ammonium thiocyanate. A liquid waveguide capillary cell (1.0 m pathlength, 550 µm i.d. and 250 µL internal volume) with a preconcentration resin were used to improve the sensitivity of the determination. Two different preconcentration resins were also tested, Chelex 100 and NTA Superflow. The developed method employing the NTA Superflow with ferrozine colorimetric reagent provided a detection limit of 0.05 µg L− 1 with a linear response up to 8 µg L− 1 and a sample throughput rate of 12 per hour. The developed system presents low reagents/sample consumptions. The accuracy was assessed using a certified reference water sample.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Microchemical Journal - Volume 93, Issue 2, November 2009, Pages 153–158
نویسندگان
, , ,