![عکس صفحه اول مقاله: On-line precipitation–dissolution in knotted reactor for thermospray flame furnace AAS for determination of ultratrace cadmium On-line precipitation–dissolution in knotted reactor for thermospray flame furnace AAS for determination of ultratrace cadmium](/preview/png/1228160.png)
A simple, environmentally friendly, cost-effective and sensitive method was developed for the determination of trace cadmium in rice and water by using flow injection (FI) on-line precipitation–dissolution in a knotted reactor (KR) as a preconcentration scheme for thermospray flame furnace atomic absorption spectrometry (TS-FF-AAS). The preconcentration was achieved by online merging the sample solution and the precipitating reagent in a KR and subsequently eluting the resultant precipitate of cadmium hydroxide with 1 mol/L HNO3. The eluant was then introduced into TS-FF-AAS for the determination. A self-assembled FI system was employed to hyphenate the KR system with TS-FF-AAS. Under optimal chemical and instrumental conditions, a limit of detection of 0.04 μg/L and a sensitivity enrichment factor of 34 for cadmium was obtained with a total initial sample volume of 4 mL. The proposed method was applied to the determination of cadmium in certified reference rice and water samples with analytical results in good agreement with their certified values. Real rice samples and real water samples were also determined by the proposed method, with analytical results confirmed by inductively coupled plasma mass spectroscopy (ICP-MS).
Journal: Microchemical Journal - Volume 91, Issue 2, March 2009, Pages 193–196