The present work set out to establish a novel stopped-flow instrument equipped with a special constructed mixing chamber containing a plunger to enable a kinetic study of the very rapid reactions under a dry inert atmosphere glove bag, in particular, for the reactions are sensitive to moisture or air. A stopped-flow spectrophotometer is essentially a conventional spectrophotometer with the addition of a system for rapid mixing of solutions. The purpose of this work is to describe the fabrication and evaluation of specially constructed and in-expensive stopped-flow system. The evaluation includes determination of the dead-time, relative mixing efficiency, and the measurement of known rate constants. Herein, a dead-time of about 3.4 ms was determined in the final modified construction of the stopped-flow apparatus in order to investigate the rapid initial during which some form of reaction intermediate is presented to be formed.
Journal: Arabian Journal of Chemistry - Volume 8, Issue 6, November 2015, Pages 873–884