کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1682889 1010485 2008 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Characterisation of a pre-cell hit detector to be used in single cell irradiation experiments at the Lund Nuclear Microprobe
موضوعات مرتبط
مهندسی و علوم پایه مهندسی مواد سطوح، پوشش‌ها و فیلم‌ها
پیش نمایش صفحه اول مقاله
Characterisation of a pre-cell hit detector to be used in single cell irradiation experiments at the Lund Nuclear Microprobe
چکیده انگلیسی
This paper describes the characterisation of an ultra-thin silicon semiconductor ΔE detector to be used as a pre-cell ion hit detector in single ion experiments on individual, living cells. The characteristics of interest for this specific application are the hit detection efficiency, which has to be close to 100% to enable bombardment with either a single ion or a counted number of ions, the beam spreading, which should be as small as possible to maintain the targeting accuracy, and the vacuum tightness, since the detector is intended, if possible, to be used simultaneously as vacuum window. The hit detection efficiency was shown to be above 99% when detecting alpha particles or 2 MeV protons, the increase in beam size was about 1 μm and the vacuum tightness was comparable to that of the Si3N4 wafer which is normally used as vacuum window, thus the ΔE detector fulfils the main criteria to function properly as a single ion hit detector.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms - Volume 266, Issue 21, November 2008, Pages 4808-4815