In analyzing Mössbauer spectra by means of the integral transmission function, a correct evaluation of the recoilless fsfs factor of the source at the position of the sample is needed. A novel method to evaluate fsfs for a 57Co source is proposed. The method uses the standard transmission experimental set up and it does not need further measurements but the ones that are mandatory in order to center the Mössbauer line and to calibrate the Mössbauer transducer. Firstly, the background counts are evaluated by collecting a standard Multi Channel Scaling (MCS) spectrum of a tick metal iron foil absorber and two Pulse Height Analysis (PHA) spectra with the same life-time and setting the maximum velocity of the transducer at the same value of the MCS spectrum. Secondly, fsfs is evaluated by fitting the collected MCS spectrum throughout the integral transmission approach. A test of the suitability of the technique is presented, too.
Journal: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms - Volume 318, Part B, 1 January 2014, Pages 253–257