کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1683687 1010511 2010 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A windowless ionization chamber for soft X-ray dosimetry
موضوعات مرتبط
مهندسی و علوم پایه مهندسی مواد سطوح، پوشش‌ها و فیلم‌ها
پیش نمایش صفحه اول مقاله
A windowless ionization chamber for soft X-ray dosimetry
چکیده انگلیسی

A new simplified model of a free air ionization chamber called a windowless air ionization chamber (WIC), for kerma in air measurements from the soft X-ray sources, has been designed and tested. The design is based on Monte Carlo calculations. The assembled WIC for testing has the electrodes in the form of half cylinders (internal diameter 36 mm; internal length 85 mm). The volume of the collecting sector of this chamber is around 1 cm3. An entrance aperture has 6 mm diameter. The results of MC calculations and tests for X-ray qualities: ranging 0.3–0.7 mm Al HVL are given.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms - Volume 268, Issue 1, 1 January 2010, Pages 92–96
نویسندگان
, , , , , ,