A gamma spectrometer including an HP Ge detector is commonly used for environmental radioactivity measurements. The efficiency of the detector should be calibrated for each geometry considered. Simulation of the calibration procedure with a validated computer program is an important auxiliary tool for environmental radioactivity laboratories. The MCNP code based on the Monte Carlo method has been applied to simulate the detection process in order to obtain spectrum peaks and determine the efficiency curve for each modelled geometry. The source used for measurements was a calibration mixed radionuclide gamma reference solution, covering a wide energy range (50–2000 keV). Two measurement geometries – Marinelli beaker and Petri boxes – as well as different materials – water, charcoal, sand – containing the source have been considered. Results obtained from the Monte Carlo model have been compared with experimental measurements in the laboratory in order to validate the model.
Journal: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms - Volume 263, Issue 1, October 2007, Pages 144–148