کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1687390 1010653 2009 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
137Cs source dose distribution using the Fricke Xylenol Gel dosimetry
موضوعات مرتبط
مهندسی و علوم پایه مهندسی مواد سطوح، پوشش‌ها و فیلم‌ها
پیش نمایش صفحه اول مقاله
137Cs source dose distribution using the Fricke Xylenol Gel dosimetry
چکیده انگلیسی

Dosimetric measurements close to radioisotope sources, such as those used in brachytherapy, require high spatial resolution to avoid incorrect results in the steep dose gradient region. In this work the Fricke Xylenol Gel dosimeter was used to obtain the spatial dose distribution. The readings from a 137Cs source were performed using two methods, visible spectrophotometer and CCD camera images. Good agreement with the Sievert summation method was found for the transversal axis dose profile within uncertainties of 4% and 5%, for the spectrophotometer and CCD camera respectively. Our results show that the dosimeter is adequate for brachytherapy dosimetry and, owing to its relatively fast and easy preparation and reading, it is recommended for quality control in brachytherapy applications.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms - Volume 267, Issue 5, March 2009, Pages 842–845
نویسندگان
, , ,