کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
179263 459343 2013 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Spectroelectrochemical measurements of redox proteins by using a simple UV/visible cell
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی مهندسی شیمی (عمومی)
پیش نمایش صفحه اول مقاله
Spectroelectrochemical measurements of redox proteins by using a simple UV/visible cell
چکیده انگلیسی


• We provide details of a UV/Vis cell for spectroelectrochemical measurements.
• This cell does not require high sample concentrations as the OTTLE cell.
• We report the spectroelectrochemical studies of various redox proteins.
• We studied cytochrome c, myoglobin, horseradish peroxidase, and cytochrome bc1.
• The measured midpoint potentials of these proteins agree with their literature values.

Details are provided of a spectroelectrochemical cell that consists of a UV/vis cuvette. We report the UV/vis spectroelectrochemical measurements of cytochrome c, myoglobin, horseradish peroxidase, and cytochrome bc1 at pH 7.0. The application on cytochrome c was performed by direct electron exchange between the protein and the working electrode. On the other hand, the measurements on myoglobin, horseradish peroxidase, and cytochrome bc1 were carried out in the presence of redox mediators. All the proteins display one-electron redox changes, as expected and measured midpoint potentials (Em) that agree with previously reported values.

Figure optionsDownload as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Electrochemistry Communications - Volume 33, August 2013, Pages 76–79
نویسندگان
, , , , , , , , ,